VV028203
【Product Name】
Manipulating Hook
【Model and Specification】
T-shaped, round elbow type, push-pull type, Y-shaped, mushroom head, slurry type, M-shaped, hemispherical type, hook type
【Product Structure and Composition】
Consisting of a head and a shaft, the head features a hook. Made of stainless steel material. Non-sterile upon delivery. Sterilization or disinfection must be performed by the user according to the instructions before use.
【Intended Use】
For hooking and pulling eye tissue.
【Contraindications】
Not for use by individuals allergic to the stainless steel used in this product.
Not for use on infected surgical sites.
【Warnings】
This product should be used by professional medical personnel only; please read all instructions before use.
Choose the appropriate model based on the intended use and follow standard surgical procedures during use.
Do not use if the product is observed to be damaged, defective, or in any other abnormal condition.
This product is provided non-sterile; thoroughly clean, dry, and sterilize the product before use.
【Precautions】
Protect the head of the product during the cleaning and sterilization processes.
Products discarded after use must be treated as medical waste according to local laws and regulations.
Avoid the use of chlorine-based cleaning agents.
Do not use a dumping method to place sharp or pointed instruments in a sterilization basin.
Avoid using solutions containing salts or bleaching agents to clean instruments or prolonged exposure to sunlight, which may cause material corrosion and rust.
Do not allow stainless steel instruments to come into contact with oxides or any acidic substances, disinfectants, salts, phenols, potassium, mercury, bleach, or other chemicals.
Products discarded after use must be treated as medical waste according to local laws and regulations.
【Usage Instructions】
Before use, check the product for any damage or missing parts.
The user should wear sterile gloves during operation.
Handle the instrument gently during use.
After use, store the instrument in an appropriate sterilization box, protecting the head.
【Sterilization Method】
The instrument should be sterilized using the hospital's standard high-temperature, high-pressure steam sterilization method before use.
【Cleaning Instructions】
Manual Cleaning: Use a specialized cleaning agent and a soft brush to clean contaminants from the instrument.
Rinsing: Rinse for at least 1 minute.
Ultrasonic Cleaning: Place the rinsed instruments in an ultrasonic cleaner for at least 10 minutes.
Rinsing: Rinse with pure or distilled water for at least 1 minute.
Drying: Place in an oven and dry at temperatures of 60-70°C for no less than 40 minutes.
【Sterilization Instructions】
Sterilize the product according to the hospital's cleaning and sterilization procedures.
-
Our company recommends high-temperature, high-pressure steam sterilization:
Standard high-temperature, high-pressure cycle: 134°C, 220 Kpa steam sterilization for 4 minutes. Other time and steam temperature cycles can also be applied. However, users must validate any deviations in time and steam temperature before use, especially as higher sterilization temperatures and pressures may damage the instrument.
Please contact your steam sterilization equipment manufacturer to confirm appropriate sterilization temperatures and times.
【Transport and Storage】
The product should be protected from heavy pressure, exposure to sunlight, and rain during transport. Store in a dry, clean environment with temperatures between 0°C and 45°C, relative humidity not exceeding 90%, and free from corrosive gases.
General transportation methods are permissible, but avoid severe impacts, vibrations, and exposure to rain or snow during transport. Handle lightly, do not overload; stacking should not exceed five layers. Packaging must ensure the product is not damaged during transport.
【Packaging】
Minimum packaging: 1 product per packaging box.
【Production Date/Expiration Date】
See label for production date. Under specified transport and storage conditions, the product can be preserved for 5 years.
【Production record certificate number】
NO. SSYJXSCB20182013
【Product record certificate number】
NO. SSXB20181020
【Product technical requirement number】
NO. SSXB20181020
【Name of registrant/manufacturer】
VisionVentions LLC
【Domicile of registrant/production enterprise】
12920 NE 125th Way Kirkland, WA USA
【Production address】
12920 NE 125th Way Kirkland, WA USA
【After-sales service unit】
VisionVentions LLC
【After-sales service address】
12920 NE 125th Way Kirkland, WA USA
【Phone】
+1 425 753 0833
【E-mail】
info@visionventions.com
【WEBSITE】
【Revision date】
January 6, 2026
【Document number】
IFU-VV028203 Rev. A